Radionuclide Therapy Waste Tank System
A waste tank system designed for controlled storage, decay and discharge of radioactive liquid waste generated in nuclear medicine and radiopharmaceutical applications.
- Glass Fiber or Stainless Steel Tank Body
- PLC or Manual Control
- Radioactive Decay Holding System
- Controlled Discharge System
Categories:
Our ProductsThe Radioactive Liquid Waste Tank System is designed to safely store radioactive wastes from nuclear medicine, radiopharmaceutical production, radiation processes and research laboratories until they decay to safe levels or are disposed of in accordance with regulatory requirements and released to the environment.
The system is not a discharge system based on radioactive dose level, but rather a decay system. The decay dose is calculated and regulated by the Turkish Nuclear Regulatory Authority (NDK) based on the facility's radioactive waste sources, considering the amount of radioactive doses received per room and the volume, and the tank retention times are also regulated accordingly.
| Tank Body | Glass fiber or stainless steel |
| Radiation Protection | The tank body can be coated with lead in areas where radiation is present. |
| Capacity | 6 m³ – 12 m³ (optional) |
| Pump System | Corrosion-resistant component pumps |
| Level Measurement | Ultrasonic level radar sensor |
| Radiation Monitoring | Room-specific radiation monitoring area |
| PLC Control Unit | Level monitoring and remote monitoring |
| Level Monitoring | Level monitoring unit |
| Connection Equipment | PPR (Polypropylene Random) main inlet and valve system |
| Emergency Safety | Overflow alarm and emergency stop circuit, bypass feature during transportation |
| Maintenance and Service | 48-hour on-site intervention, spare parts and maintenance provided |
| Decontamination Management | Tank cleaning calculated by NDK |
| Discharge System | Discharge line based on controlled analysis and monitoring |